features |
|
|
| |
›› calibrated in ml/hr
›› 50ml to microliter syringes
›› flow rates from0 .025ml/hr to 430ml/hr
›› amluracy within 2%
›› on/off lighted rocker switch
›› stainless steel used for all critical parts |
|
|
|
|
model A99-E
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country)
|
$689
$719
|
| Infusion rates from 0.25ml/hr to 143ml/hr with syringe
sizes from 5ml to 50ml. The infusion rate is set on a 2 digit counter.
A stop switch disconnects power at the end of the syringe. |
|
 |
|
|
|
| |
|
|
| |
base
model with options |
|
|
|
|
|
|
|
|
| |
|
|
|
|
| |
model
A99-EJ
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country) |
$723
$741
|
|
“J” option will allow the flow rate to be increased by 3 times
(0.25ml/hr to 430ml/hr)iv>
|
| |
model
A99-EM
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country) |
$723
$741
|
|
“M” option will allow the flow rate to be decreased to .10 the
normal speed (0.025ml/hr to 143ml/hr)
|
| |
model A99-EJM
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country) |
$741
$782
|
|
Infusion rates from 0.025ml/hr to 430ml/hr with syringe sizes from 5ml
to 50ml. The infusion rate is set on a four digit selector with 299.9
rates. A stop switch disconnects power when the plunger reaches the
end of the syringe. |
| |
|
|
|
|
| |
add-on options |
|
|
|
|
|
|
|
|
| |
option B
|
$45 |
|
Reversing Switch for withdrawal |
| |
option F
|
$76 |
|
Alarm option |
| |
option W |
Call |
|
Key interlock for IV pole plate |
| |
Aml-BP
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country) |
$264
$313 |
|
Battery Pack
auxiliary power for A99-E—12 Volts,
7 Ampere Hours |
| |
Aml-99 M
|
$83 |
|
Replacement motor
for A99-E pumps |
| |
Aml-IV Pole
Plate |
Call |
|
Mounts A99-E to
IV pole |
| |
|
|
|
|
| |
specifications |
|
|
|
|
| |
FLOW RATES
model A99-E |
The Model A99-E can accommodate from a 50ml syringe to a
microliter syringe, both in plastic and in glass. The Model
A99-E has a built in flow chart, calibrated for the Multifit
brand syringe, with 99 equally spaced speeds from:
1.45 to 143ml/hr with 50ml syringe
0.95 to 94ml/hr with 3Oml syringe
0.71 to 70ml/hr with 20ml syringe
0.40 to 39ml/hr with 10ml syringe
0.25 to 25ml/hr with 5ml syringe
To determine flow rates with other syringes, a calibration
sheet is included with every pump.
|
FLOW RATES
model A99-EJM |
The Model A99-EJM can accommodate from a
50ml syringe to a microliter syringe, both in plastic and
in glass. The Model A99-EJM has a built in flow chart,
calibrated for the Multifit brand syringe, with 299.9
equally spaced speeds from:
.14 to 430ml/hr with 50ml syringe
.09 to 282ml/hr with 30ml syringe
.07 to 210ml/hr with 20ml syringe
.04 to 117ml/hr with 10ml syringe
.025 to 75ml/hr with 5ml syringe
To determine flow rates with other syringes, a calibration
sheet is included with every pump.
|
FLOW RATES
model A99-EJ |
The Model A99-EJ can accommodate from a 50ml syringe to a microliter syringe, both in
plastic and in glass. The Model A99-EJ has a built in
flow chart, calibrated for the Multifit brand syringe,
with 300 equally spaced speeds from:
1.45 to 430ml/hr with 50ml syringe
0.95 to 282ml/hr with 30ml syringe
0.71 to 210ml/hr with 20ml syringe
0.40 to 117ml/hr with 10ml syringe
0.25 to 75ml/hr with 5ml syringe
To determine flow rates with other syringes, a
calibration sheet is included with every pump.
|
FLOW RATES
model A99-EM |
The Model A99-EM can accommodate from a 50ml syringe to a microliter syringe, both in
plastic and in glass. The Model A99-EM has a built
in flow chart, calibrated for the Multifit brand
syringe, with 99.9 equally spaced speeds from:
.145 to 143ml/hr with 50ml syringe
.095 to 94ml/hr with 30ml syringe
.071 to 70ml/hr with 20ml syringe
.040 to 39ml/hr with 10ml syringe
.025 to 25ml/hr with 5ml syringe
To determine flow rates with other syringes, a
calibration sheet is included with every pump.
|
|
All critical parts are
fabricated from stainless steel, and bronze
bearings are lubricated for life. Thus a minimum
of maintenance is required. A lead screw, which is
directly coupled to a synchronous motor eliminates
the need for expensive gear trains, or inamlurate
friction drives. A clutch built into the
instrument is designed to prevent the motor from
burning out.
|
|
Accuracy:
Overall accuracy within 2%, repeatability within
.5%. Calibrated at 115 volts A.C. Less than 1%
change in
flow with a 15 volt difference.
Pressure limits: Accurate flow up to 5 psi or 270 mm/hg with
a
50-60ml syringe; 7 psi or 400 mm/hg with a 30-35ml
syringe; 10 psi or 500 mm/hg with 20ml syringe;
20 psi or 1000 mm/hg with 10ml syringe.
Electrical: 3 wire grounded plug, 15 watts @ 115 volts 50-60
hertz. Less than 25 microamperes current
leakage.
Physical: Weight—7.5 pounds. Size- 6" high x 6" wide x 10"
long.
|
|
| |
| |
|
|
|
|
|