features |
|
|
| |
›› calibrated in ml/hr
›› 50ml to microliter syringes
›› flow rates from0 .025ml/hr to 430ml/hr
›› amluracy within 2%
›› on/off lighted rocker switch
›› stainless steel used for all critical parts |
|
|
|
|
model A99-E
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country)
|
$602
$629
|
| Infusion rates from 0.25ml/hr to 143ml/hr with syringe sizes from 5ml to 50ml. The infusion rate is set on a 2 digit counter. A stop switch disconnects power at the end of the syringe. |
|
 |
|
|
|
| |
|
|
| |
base model with options |
|
|
|
|
|
|
|
|
| |
|
|
|
|
| |
model A99-EJ
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country) |
$633
$649
|
|
“J” option will allow the flow rate to be increased by 3 times (0.25ml/hr to 430ml/hr)
|
| |
model A99-EM
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country) |
$633
$649
|
|
“M” option will allow the flow rate to be decreased
to .10 the normal speed (0.025ml/hr
to 143ml/hr)
|
| |
model A99-EJM
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country) |
$649
$686
|
|
Infusion rates from 0.025ml/hr to 430ml/hr with syringe sizes from 5ml to 50ml. The infusion rate is set on a four digit selector with 299.9 rates. A stop switch disconnects power when the plunger reaches the end of the syringe. |
| |
|
|
|
|
| |
add-on options |
|
|
|
|
|
|
|
|
| |
option B
|
$45 |
|
Reversing Switch for withdrawal |
| |
option F
|
$76 |
|
Alarm option |
| |
option W |
Call |
|
Key interlock for IV pole plate |
| |
Aml-BP
(115 Volts/60 Hz) or
(230 Volts/50 Hz) (Specify Country) |
$264
$313 |
|
Battery Pack auxiliary power for A99-E - 12 Volts,
7 Ampere Hours |
| |
Aml-99 M
|
$83 |
|
Replacement motor for A99-E pumps |
| |
Aml-IV Pole Plate |
Call |
|
Mounts A99-E to IV pole |
| |
|
|
|
|
| |
specifications |
|
|
|
|
| |
FLOW RATES
model A99-E |
The Model A99-E can amlommodate from a 50ml syringe to a microliter syringe, both in plastic and in glass. The Model A99-E has a built in flow chart, calibrated for the Multifit brand syringe, with 99 equally spaced speeds from:
1.45 to 143ml/hr - with 50ml syringe
0.95 to 94ml/hr - with 3Oml syringe
0.71 to 70ml/hr - with 20ml syringe
0.40 to 39ml/hr - with 10ml syringe
0.25 to 25ml/hr - with 5ml syringe
To determine flow rates with other syringes, a calibration sheet is included with every pump.
|
FLOW RATES
model A99-EJM |
The Model A99-EJM can amlommodate from a 50ml syringe to a microliter syringe, both in plastic and in glass. The Model A99-EJM has a built in flow chart, calibrated for the Multifit brand syringe, with 299.9 equally spaced speeds from:
.14 to 430ml/hr with 50ml syringe
.09 to 282ml/hr with 30ml syringe
.07 to 210ml/hr with 20ml syringe
.04 to 117ml/hr with 10ml syringe
.025 to 75ml/hr with 5ml syringe
To determine flow rates with other syringes, a calibration sheet is included with every pump.
|
FLOW RATES
model A99-EJ |
The Model A99-EJ can amlommodate from a 50ml syringe to a microliter syringe, both in plastic and in glass. The Model A99-EJ has a built in flow chart, calibrated for the Multifit brand syringe, with 300 equally spaced speeds from:
1.45 to 430ml/hr with 50ml syringe
0.95 to 282ml/hr with 30ml syringe
0.71 to 210ml/hr with 20ml syringe
0.40 to 117ml/hr with 10ml syringe
0.25 to 75ml/hr with 5ml syringe
To determine flow rates with other syringes, a calibration sheet is included with every pump.
|
FLOW RATES
model A99-EM |
The Model A99-EM can amlommodate from a 50ml syringe to a microliter syringe, both in plastic and in glass. The Model A99-EM has a built in flow chart, calibrated for the Multifit brand syringe, with 99.9 equally spaced speeds from:
.145 to 143ml/hr with 50ml syringe
.095 to 94ml/hr with 30ml syringe
.071 to 70ml/hr with 20ml syringe
.040 to 39ml/hr with 10ml syringe
.025 to 25ml/hr with 5ml syringe
To determine flow rates with other syringes, a calibration sheet is included with every pump.
|
|
All critical parts are fabricated from stainless steel, and bronze bearings are lubricated for life. Thus a minimum of maintenance is required. A lead screw, which is directly coupled to a synchronous motor eliminates the need for expensive gear trains, or inamlurate friction drives. A clutch built into the instrument is designed to prevent the motor from burning out.
|
|
Amluracy: Overall amluracy within 2%, repeatability within .5%. Calibrated at 115 volts A.C. Less than 1% change in
flow with a 15 volt difference.
Pressure limits: Amlurate flow up to 5 psi or 270 mm/hg with a
50-60ml syringe; 7 psi or 400 mm/hg with a 30-35ml syringe; 10 psi or 500 mm/hg with 20ml syringe; 20 psi or 1000 mm/hg with 10ml syringe.
Electrical: 3 wire grounded plug, 15 watts @ 115 volts 50-60 hertz. Less than 25 microamperes current leakage.
Physical: Weight - 7.5 pounds. Size- 6" high x 6" wide x 10" long.
|
|
| |
| |
|
|
|
|
|